GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
Viewer
11-viewer.fm Page no. 117 Chapter 11
Repeated or Rejected Images
Images qualified as Repeated or Rejected are not deleted from the AWS database.
Even in the specific case with:
- Auto Push Repeated images? and Auto Push Rejected images? both set to NO in Medical Appli-
cation preferences, Quality Check part,
- Storage commitment activated on the station,
- Auto delete: ON,
the Repeated and Rejected images are not deleted unless they are manually pushed to a host.
4-5-1-2 Procedure to qualify an image
1. While viewing a processed image in acquisition mode, the QUALITY CHECK OK button in the Viewer
Function panel indicates the current level (QUALITY CHECK OK or QUALITY CHECK NOK): default
level is QUALITY CHECK OK.
If you decide that the quality of an image is not satisfactory, click the QUALITY CHECK OK button to
open the Repeat and Reject Selection window:
Display Parameters
Inv..
Standard
Reprocess
QUALITY
CHECK OK
Repeat and Reject Selection
1 - Mark the selected image as Accepted, Repeated or Rejected
Accepted
Repeated (must do the exposure again to complete exam)
Rejected (no clinical value, no additional exposure to complete exam)
2 - Select a Repeat or Reject reason
Positioning
Patient Motion
Poor Compression
Improper Detector Exposure
X-Ray Equipment Failure
Equipment Artifacts
Blank Image
Clinical Artifacts
Incorrect View Marker
QC, Acceptance Tests, Calibration
Interventional Image (e.g., wire loc.)
Other
If "Other" is selected, please provide a reason in the text box below
Mark each image already acquired to status and reason selected
Save
Cancel
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals