GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
Error messages
16-errors.fm Page no. 187 Chapter 16
H01 Button interface
comm. failure
The system is set in emergency
mode:
- A button failure occurred;; the
system is set in "halt" state.
- It is not possible to perform any
movement or image acquisition.
1) Decompress the patient using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H02 Compression
comm. failure
The system is set in emergency
mode:
- The compression device is not
communicating; the system is set
in "halt" state.
- It is not possible to perform any
movement or image acquisition.
1) Decompress the patient using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H03 Compression con-
trol failure
The system is set in emergency
mode:
- The compression device has
failed; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H04 Internal software
error
The system is set in emergency
mode:
- An error occurred on the Gantry
CPU; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H05 Internal software
error
The system is set in emergency
mode:
- An error occurred on the Gantry
CPU; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H06 Lift communica-
tion failure
The system is set in emergency
mode:
- The Lift device is not communi-
cating, the system is set in "halt"
state.
- It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H07 Lift control failure The system is set in emergency
mode:
- The Lift device is failed; the sys-
tem is set in "halt" state.
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
Message on
Gantry Readout
Explanation Corrective Action