Chapter 16 Page no. 188 16-errors.fm
GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
Error messages
H08 Power supply unit
comm. Failure
The system is set in emergency
mode:
- The Power unit is not communi-
cating; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H09 Reset the system.
Reload the SW
The system is set in emergency
mode:
- An error occurred on the Gantry
CPU; the system is set in "halt"
state. (occurred only during soft-
ware upgrade)
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H10 Reset the system.
Reload the SW
The system is set in emergency
mode:
- An error occurred on the Gantry
CPU; the system is set in "halt"
state. (occurred only during soft-
ware upgrade)
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H11 Reset the system.
Reload the SW
The system is set in emergency
mode:
- An error occurred on the Gantry
CPU; the system is set in "halt"
state. (occurred only during soft-
ware upgrade)
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H12 Reset the system.
Reload the SW
The system is set in emergency
mode:
- An error occurred on the Gantry
CPU; the system is set in "halt"
state. (occurred only during soft-
ware upgrade)
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H13 Rotation communi-
cation failure
The system is set in emergency
mode:
- The Rotation device is not com-
municating, the system is set in
"halt" state.
- It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
Message on
Gantry Readout
Explanation Corrective Action
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).