Variable Therapy Mode
In Variable Therapy mode, the high pressure time (Up Time) and low pressure time (Down Time) will
also be displayed on the main screen. The user will be asked to set 5 parameters when selecting:
1. Treatment Time. Treatment time allows the user to set how long they want the patient to be on
Variable Therapy mode. Once the treatment time ended, the system will automatically switch
back to Continuous Therapy mode.
2. Up Time. Up time allows the user to determine how long they want the system to hold at a set
high pressure vacuum. When the time is up, it will go down to the set down pressure and will
remain at that level until the down time ends. The whole process will then cycle up and down
until the treatment time finishes.
3. Up Pressure. Up pressure allows the user to determine the high vacuum threshold while the
patient is on Variable Therapy.
4. Down Time. Down time allows the user to determine how long they want the system to hold at
a set low pressure vacuum. When the time is up, it will go up to the set up pressure and will
remain at that level until the up time ends. The whole process will then cycle down and up until
the treatment time finishes.
5. Down Pressure. Down pressure allows the user to determine the low vacuum threshold while
the patient is on Variable Therapy.