Electromagnetic Compatibility
The Genadyne XLR8+ is intended for use in the electromagnetic environment specified below. The customer or the user of the Genadyne
XLR8+ should assure that it is used in such an environment.
Immunity Test
IEC 60601 test level
Compliance level
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
The Genadyne XLR8+ is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Genadyne XLR8+ can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Genadyne XLR8+ as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12 m
0.12 m
0.23 m
0.1
0.37 m
0.37 m
0.74 m
1
1.17 m
1.17 m
2.33 m
10
3.69 m
3.69 m
7.38 m
100
11.67 m
11.67 m
23.33 m