Genadyne Biotechnologies | DMR-06-091- Rev F
Electromagnetic Compatibility
The Genadyne XLR8+ is intended for use in the electromagnetic environment specified below. The customer or the user of the Genadyne
XLR8+ should assure that it is used in such an environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
The Genadyne XLR8+ is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Genadyne XLR8+ can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Genadyne XLR8+ as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter (m)