Genadyne Biotechnologies | DMR-06-091- Rev F
31
Electromagnetic Compatibility
Electromagnetic Compatibility
The Genadyne XLR8+ is intended for use in the electromagnetic environment specified below. The customer or the user of the Genadyne
XLR8+ should assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic environment
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
Passed
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast transient /
burst
IEC 61000-4-4
+/- 2 kV for power supply lines
+/- 1 kV for input / output lines
Below
Maximum
permissible limit
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
Acceptable
Performance
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
< 5% Ut
(>95 % dip in Ut) for 0.5 cycle
40% Ut
(60% dip in Ut) for 5 cycles
70% Ut (30% dip in Ut) for 25
cycles
<5% Ut
(>95% dip in Ut) for 5 sec
Acceptable
Performance
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the Genadyne XLR8+
be powered from an
uninterruptable power supply or
a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
Non Applicable
Power frequency magnetic
fields should be at levels
characteristics of a typical
location in a typical commercial
or hospital environment
Note Ut is the a.c. mains voltage prior to application of the test level
Guidance and manufacturer’s declaration – electromagnetic emissions
The Genadyne XLR8+ is intended for use in the electromagnetic environment specified below. The
customer or the user of the Genadyne XLR8+ should assure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Genadyne XLR8+ uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The Genadyne XLR8+ is suitable for use in all establishments
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies