Genadyne Biotechnologies | DMR-06-091- Rev F
Therapy Selection
To select which therapy to use at any time:
1. Press the Menu / Select button.
2. Scroll using the Up button or Down button and choose the Treatment Mode function by pressing
the Menu / Select button once.
3. Choose either Continuous or Variable Therapy by pressing the Menu / Select button once.
4. For Continuous Therapy selection, after Step 3, exit to the main screen by holding on to the Exit
/ Cancel button for 5 seconds. The user can also press the Exit / Cancel button 2 times or more
to exit to the main screen.
5. For Variable Therapy selection, after Step 3, press Menu / Select button one more time to enter
into the variable setting screen.
a. Treatment Time. Press the Menu / Select button to enter the desired treatment time. For
continuous variable mode, set the treatment time to 0h. Use the Up button or Down button
to increase or decrease the desired time. All settings are in hours. Once the treatment
time is set, press the Menu / Select button again to confirm selection. It will then bring you
back to the Variable Therapy setting screen.
b. 125/30. 5 Minutes at 125 mmhg, then 3 minutes at 30 mmhg.
c. 80/30. 5 Minutes at 80 mmhg, then 3 minutes at 30 mmhg.
d. 125/0. 5 Minutes at 125 mmhg, then 3 minutes at 0 mmhg.
e. Program. Press the Menu / Select button to define your own variable therapy session.
i. Up Time. Press the Menu / Select button to enter the desired up time. Use the Up
button or Down button to increase or decrease the desired time. All settings are in
minutes. Once the Up Time is set, press the Menu / Select button again to confirm
selection. It will then bring you back to the Variable Therapy setting screen.
ii. Up Pressure. Press the Menu / Select button to enter the desired high pressure
threshold. Use the Up button or Down button to increase or decrease the desired
vacuum pressure. All settings are in mmHg. Once the vacuum pressure is set,
press the Menu / Select button to confirm selection. It will then bring you back to
the Variable Therapy setting screen.
iii. Down Time. Press the Menu / Select button to enter the desired down time. Use
the Up button or Down button to increase or decrease the desired time. All settings
are in minutes. Once the Down Time is set, press the Menu / Select button to
confirm selection. It will then bring you back to the Variable Therapy setting screen.
iv. Down Pressure. Press the Menu / Select button to enter the desired low pressure
threshold. Use the Up button or Down button to increase or decrease the desired
vacuum pressure. All settings are in mmHg. Once the vacuum pressure is set,
press the Menu / Select button to confirm selection. It will then bring you back to
the Variable Therapy setting screen.
6. To exit the variable setting screen and return to the main screen, hold on to the Exit / Cancel
button for 5 seconds. The user can also press the Exit / Cancel button 3 times or more to exit to
the main screen.