Maintenance
It is mandatory for the product to have scheduled diagnostic maintenance every 12 months or 6000
hours, whichever comes first. Failure to comply will void warranty.
By following the steps below, the user can determine that the unit is due for maintenance:
1. Press Menu / Select button
2. Navigate to Advanced Menu by using the Up /Down button and press the Menu / Select button
3. Navigate to the System Info by using the Up / Down button and press the Menu / Select button
4. If Preventive Maintenance is required, the below can be found in the System Info Menu
a. MAINTENANCE REQ. : YES
Product is also recommended to be opened and inspected between patient use by trained personnel
to detect any possible issues before returning the device to the field. Please contact Genadyne for a
free training on how to perform both the recommended and mandatory preventative maintenance on
the device.
The mandatory preventative maintenance consists of full inspection and diagnostics of unit,
replacement of internal tubing and battery if necessary, replacement of Double O-rings, replacement
of odor patches and cleaning of the inside and outside of the unit. This maintenance insures proper
continued performance from the unit, as well as maintaining the indicated battery run time.
Please contact your local distributor or Genadyne regarding the preventative maintenance for the
device.
Cleaning
Adherence to facility directives concerning hygiene is of prime importance.
Only use low level diluted form of disinfectants or cleaning agents when cleaning the XLR8+. Use
damped cloth to clean the pump.
Be cautious when cleaning because no liquids should enter the power unit. If liquid goes inside of the
power unit, it might cause the unit to malfunction or damage the mechanics.
Dry with a separate soft cloth.
Do not use solvents or abrasives.
Do not immerse any part of the XLR8+ in fluid or use an unnecessarily wet cloth.
Please contact your distributor if any liquids penetrated the device.