Manually-operated door
When the door is closed manually and V is pressed, the door is locked automatically,
the yellow lamp ashes and the program starts.
When the program is complete, a green lamp lights up on both sides (if there are double
doors) and the clean-side door is unlocked. When the door has been opened manually,
the green lamp goes out. With double doors, the clean-side door must be closed and
locked by pressing the W button before the soiled-side door can be unlocked.
Automatically-operated door
When V is pressed, the door is locked, the yellow lamp ashes and the program starts.
When the program is complete, a green lamp lights up on both sides (if there are double
doors) and the clean-side door opens automatically. When the door is fully open the green
lamp goes out on both sides. With double doors, the clean-side door must be closed and
locked by pressing the W button before the soiled-side door opens automatically.