<Doc_USR><Doc_502404300><Rel_A><Lang_GB>
Daily inspection
Check that no objects have fallen down into the bottom of the washing chamber.
Loose objects may cause blockages and seriously reduce the performance of the
washer disinfector.
Check that the spray arms can rotate freely. Check that the holes in the spray arms
are not blocked. Blocked spray holes may adversely affect the cleaning result.
Check that the goods to be cleaned are retained in the insert.
If the arms need cleaning, remove them by unscrewing the central nut. The arm
can then be lifted off and washed clean. Check that the arm can rotate freely when
re-installed.
Remove and clean the strainer in the bottom of the washing chamber. A dirty
coarse strainer may prevent water from circulating and create the conditions for the
growth of bacteria.
If this is required for cleaning, open the strainer by removing the spray arm. To do
this, slacken the locknut and move it sideways. Then, using the handles, turn the
counter-clockwise and lift it out.
Open the detergent containers and check that they contain detergent/rinse-aid.
Check that the wash trolley docks properly to the water connections.
Check the accessories of the machine visually with regard to their intended use and
function. For example, blocked ducts, worn and missing parts that might affect the
performance of the machine. Water must be able to ow freely through the ducts in
tubular instruments.
Check that the spring movement of the docking mechanisms moves easily. Press in
the docking mechanism and release it. It must return to its initial position.
Check the cleaning instructions of the goods manufacturer.
To achieve the intended level of performance and safety, distribute the load evenly
over the washing surface.
Check the cleaning result visually when the process is complete.
The spray arm journal must be checked daily and replaced every ve years.
Check the process valve for deposits and dirt in the diaphragm.
Check the drain valve for deposits and dirt in the diaphragm.
Spray arm
Coarse strainer
V2025b
Locknut

Other manuals for Getinge 88-SERIES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 88-SERIES and is the answer not in the manual?

Getinge 88-SERIES Specifications

General IconGeneral
BrandGetinge
Model88-SERIES
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals