OPERATING INSTRUCTIONS
Machine without barcode scanner
Loading goods to be washed into the machine
Load the items to be washed on to the trolley and run the trolley into the machine.
Examples showing how to load the trolley are shown on the goods positioning sign
and in the loading handbook.
The trolley may be loaded with Max 75 kg.
Make sure that the washer arms can rotate freely without touching the goods.
Select the program.
Choose a program in one of the following ways:
- Press one of the program selection buttons 1 -6
- Press S. The display now shows “CHOOSE PROGRAM”.
Press S again.
Choose a program from the list with H and J.
Press S to conrm your choice.
You can see the machine programs on the separate program sheet.
Press V. A yellow lamp ashes for 5 seconds. During this time, the program can
be aborted by pressing V again.
A fine filter must be used when cleaning and disinfecting tubular instruments
with a passage smaller than 2 mm, to prevent particles getting stuck in the
instruments. Blocked or reduced water flow may result in cleaning and disin-
fection failing to take place.
If the program is cancelled after start, the cleaning result and disinfection will
be unsuccessful.
While a program is running
While the program is running, a yellow lamp at M is lit and the program phase (e.g.
“Rinse”) and the current temperature are shown on the display. Press S to see how
much time remains.
When the program is complete
When the program is complete, a green lamp lights up at N and the clean-side
door opens automatically.
If the items are still hot after a program, handle them with care to avoid burns.
Empty the machine, close and lock the clean-side door by pressing W.
If the same program is to be used again, there is no need to select the program
again. Simply press V once.
Note!
The display switches to standby mode after 30 minutes and returns automatically when a
program is started or selected.

Other manuals for Getinge 88-SERIES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 88-SERIES and is the answer not in the manual?

Getinge 88-SERIES Specifications

General IconGeneral
BrandGetinge
Model88-SERIES
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals