Maintenance and inspection procedures
Mechanical inspections
24 / 44
Lucea 50_100
01740 EN 01
Fig.27: Integrity of the lightheads
Integrity of the lightheads, brake screw cap
and mounting screws
1. Check the lightheads for chipped paint, im-
pact marks, any other damage, loose cov-
ers, etc.
2. Check that the cap protecting the brake
screw is properly seated.
3. Check that all the mounting screws are
present.
4. If a problem is noted, contact technical
support.
Fig.28: DF fork plugs
Fork plugs (DF version only)
1. Check that the grey plugs are correctly in-
stalled.
2. If a problem is noted, contact technical
support.
For the mobile version:
Fig.29: Device verticality
Check the verticality and stability of the
device.