EasyManua.ls Logo

GGM VH-2600A-230 - 3.4. Display Symbols

Default Icon
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
VH-2600A
1. Locked / Unlock
*System will automatic “Lock” the panel after 60 sec when power on, and display the
chamber temperature.
*Hold FUNCTION for 3 sec > Unlock
*Hold FUNCTION for 3 sec > Lock
The system will relock automatically after 60 sec, if there is no more operation.
2. Exchange Temperature
1. Heater Wire Mode:
The system will automatically swift to “Heater Wire Mode”, when the system equipped
with heater wire while power on. System will swift to “Non-
Heater Wire Mode” after
holding MODE 3 sec.
2. Patient Mode
*In Patient Mode, button will be locked after 60 sec when power on.
*After power on, it is in “Invasive” Mode.
*Hold FUNCTION for 3 sec + Push MODE1 times and system will swift to
“Non-Invasive” Mode.
Will silent all alarm sound for 3 min, the alarm indicator will keep on flashing until the
alarm situation improved.
Invasive mode
Heater Wire ModeAirway 40°CChamber 37°C
Non-Heater Wire ModeAirway 37°CChamber66°C
Non-Invasive mode
Heater Wire ModeAirway 34°CChamber 31°C
Non-Heater Wire ModeAirway 31°CChamber66°C

Table of Contents

Related product manuals