VH-2600A
1. Locked / Unlock
*System will automatic “Lock” the panel after 60 sec when power on, and display the
chamber temperature.
*Hold FUNCTION for 3 sec => Unlock
*Hold FUNCTION for 3 sec => Lock
【The system will relock automatically after 60 sec, if there is no more operation.】
2. Exchange Temperature
1. Heater Wire Mode:
The system will automatically swift to “Heater Wire Mode”, when the system equipped
with heater wire while power on. System will swift to “Non-
Heater Wire Mode” after
holding MODE 3 sec.
2. Patient Mode
*In Patient Mode, button will be locked after 60 sec when power on.
*After power on, it is in “Invasive” Mode.
*Hold FUNCTION for 3 sec + Push MODE1 times and system will swift to
“Non-Invasive” Mode.
Will silent all alarm sound for 3 min, the alarm indicator will keep on flashing until the
alarm situation improved.
Invasive mode
Heater Wire Mode:Airway 40°C;Chamber 37°C。
Non-Heater Wire Mode:Airway 37°C;Chamber≦66°C。
Non-Invasive mode
Heater Wire Mode:Airway 34°C;Chamber 31°C。
Non-Heater Wire Mode:Airway 31°C;Chamber≦66°C。