EasyManua.ls Logo

GGM VH-2600A-230 - 5. Alarm and Troubleshooting; 5.1. Alarm Indicator

Default Icon
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14
Wipe the humidifier clear of any cleaning residues before use.
1. Disconnect the humidifier from any electrical outlet.
2. Clean the humidifier with one of the following recommended disinfectants using a damp cloth:
a. Isopropyl Alcohol
b. Normal dishwashing detergent
NOTE: Follow the manufacturers instructions carefully.
NOTE: Use the correct dilution of the disinfectant.
NOTE: DO NOT immerse the humidifier in any liquid.
3. Wipe the humidifier clear of any cleaning residues before use.
4.1.2. Temperature Probe
The cleaning methods listed below comply with FDA regulations for these types of equipment and do not
affect the integrity or performance of the temperature probe.
Please contact your local dealer for suggestions on other cleaning methods not mentioned here.
1. Clean the probe and remove all visible contaminants by physically.
2. Use the following solutions to sterilize the temperature probeSporicidin
TM
Sporox
TM
Cidex
TM
Cidex
TM
OPA or Ethylene Oxide sterilization. The temperature probe was sterilized at 55 ° C (131 ° F) and 80 kPa,
allowing the remaining ETO to evaporate at least 15 hours before use.
NOTEPlease follow the manufacturer's instructions. Use the correct dilution disinfectant.
WarningDo not immerse the connector plug in the disinfectant.
Do not use autoclaves.
Do not use detergent or solvent.
3. Wipe the patient end temperature probe before use
4. Stored in a clean environment.
4.1.3. Heater Wire Adapter
Clean the heater wire at with one of the following recommended disinfectants using a damp cloth:
a. Isopropyl Alcohol
b. Normal dishwashing detergent
4.1.4. Mains Plug
If the mains plug detachable design, check whether the screw is locked or not.

Table of Contents

Related product manuals