EasyManua.ls Logo

Gilson pipetman - 18 SPECIFICATIONS

Gilson pipetman
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 26 -
18 SPECIFICATIONS
PIPETMAN M is a high quality pipette that of-
fersexcellentaccuracyandprecision;itisfully
compliantwithISO8655andisCEmarked(for
IVDandEMCdirectives).
The gures given in theGilson Maximum
Permissible Errors table were obtained
using “PIPETMAN TIPS.Theseguresareonly
guaranteed when using Genuine PIPETMAN
TIPS.
Each pipette is inspected and validated
by qualied technicians according to the
Gilson Quality System. Gilson declares that
its manufactured pipettes comply with the
requirements of the ISO 8655 Standard, by
type testing. The adjustment is carried out
understrictlydenedandmonitoredcondi-
tions(ISO8655-6):
• Basisofadjustment,Ex.
• Referencetemperature,20°C
• Relativehumidity,50%
• Barometricpressure,101kPa
• Useofdistilledwatergrade3(ISO3696)
• Ten measurements for each test volume,
whichareNominalVolume,50%ofNominal
Volumeand10%ofNominalVolume.
(tested mode: PIPET, speed 6 with PIPETMAN Tips)
Maximum Permissible Errors
Model PIPETMAN Volume Gilson ISO 8655
(Reference) tips used Systematic Random Systematic Random
error (L) error (L) error (L) error (L)
P8x20M
Min. 1 ± 0.08 ≤ 0.05 ± 0.4 ≤ 0.2
(F81005) DL10
P12x20M
10 ± 0.15 ≤ 0.10 ± 0.4 ≤ 0.2
(F81006)
Max. 20 ± 0.25 ≤ 0.12 ± 0.4 ≤ 0.2
P8x200M
Min. 20 ± 0.50 ≤ 0.16 ± 3.2 ≤ 1.2
(F81007)
D200
P12x200M
100 ± 1.00 ≤ 0.30 ± 3.2 ≤ 1.2
(F81008)
Max. 200 ± 2.00 ≤ 0.50 ± 3.2 ≤ 1.2
P8x300M Min. 10 ± 1.00 ≤ 0.18 ± 8.0 ≤ 3.0
(F81011)
D300
30 ± 1.00 ≤ 0.18 ± 8.0 ≤ 3.0
P12x300M 150 ± 1.50 ≤ 0.375 ± 8.0 ≤ 3.0
(F81012) Max. 300 ± 2.40 ≤ 0.45 ± 8.0 ≤ 3.0
Maximum Permissible Errors
Model PIPETMAN Volume Gilson ISO 8655
(Reference) tips used Systematic Random Systematic Random
error (L) error (L) error (L) error (L)
P10M Min. 0.5 ± 0.04 ≤ 0.013 ± 0.12 ≤ 0.08
(F81001) DL10 1 ± 0.025 ≤ 0.012 ± 0.12 ≤ 0.08
5 ± 0.060 ≤ 0.020 ± 0.12 ≤ 0.08
Max. 10 ± 0.080 ≤ 0.025 ± 0.12 ≤ 0.08
P20M Min. 2 ± 0.075 ≤ 0.025 ± 0.20 ≤ 0.10
(F81002) D200 10 ± 0.100 ≤ 0.035 ± 0.20 ≤ 0.10
Max. 20 ± 0.150 ≤ 0.05 ± 0.20 ≤ 0.10
P200M Min. 20 ± 0.40 ≤ 0.15 ± 1.60 ≤ 0.60
(F81003) D300 100 ± 0.80 ≤ 0.22 ± 1.60 ≤ 0.60
Max. 200 ± 1.00 ≤ 0.26 ± 1.60 ≤ 0.60
P1000M Min. 100 ± 2.50 ≤ 0.4 ± 8 ≤ 3.0
(F81004) D1000 500 ± 3.50 ≤ 0.7 ± 8 ≤ 3.0
Max. 1000 ± 5.00 ≤ 1.1 ± 8 ≤ 3.0

Other manuals for Gilson pipetman