2-5
Installation 2
Electrical Setup
Contact Connections
The VALVEMATE contains two input contact
pairs and a removable terminal block connector for
easy wiring. The inputs are labeled for quick
identification.
Input contacts
Via the input contacts, the VALVEMATE can
receive signals to turn the valve in a clockwise
(CW) or counterclockwise (CCW) direction.
When controlling the VALVEMATE via the input
contacts, send pulse signals. These signals
momentarily close (activate) and open (deactivate)
a contact. The duration of the pulse must be at least
0.1 second.
Items you will need
To make connections to the input and output
contacts, you need:
a wire insulation stripper
a small blade screwdriver
a 2-conductor cable for each connection that
you will make
You can purchase a 6-foot piece of suitable cable
(part number 709910206) from Gilson.