3-3
Operation 3
Front Panel Control
5 Up and down In position mode, use these
arrow keys keys to rotate the valve. For a
multi-position valve, pressing
the up arrow key turns the
valve in a counterclockwise
direction and pressing the
down arrow key turns it in a
clockwise direction. For a 2-
position valve, pressing an
arrow key turns the valve to
its other position.
In unit ID mode, use these
keys to select a unit ID from
30 to 39. The default is 35.
You need to change the
VALVEMATE’s unit ID if it
and another VALVEMATE
are being externally controlled
via their GSIOC or RS-232
ports.
In baud rate mode, use these
keys to select among the
following baud rates: 9600,
19200, or External.
In valve mode, use these keys
to identify the installed valve.
See Appendix A for a list of
valves.
In the service mode, these
keys are not used.