5-4
Troubleshooting 5
Troubleshooting
Fluid flowing from wrong port or not flowing
from any port
The valve may be misaligned; turn power OFF
then ON to the VALVEMATE
The valve may not have been in home position
when installed in the VALVEMATE; remove
the valve and use an adjustable wrench to place
it in home position then reinstall the valve (refer
to Section 2)
The wrong valve adapter may be attached to
the valve; remove the valve and verify that the
correct adapter is attached (refer to Section 2)
The valve may be incorrectly plumbed; refer to
the instruction sheet supplied with the valve
Fluid may not be reaching the valve; if
connected to a pumping system, lower system
pressure and disconnect the valve’s inlet tubing
to determine if fluid is reaching the valve
Low flowrate
Check for leaks in tubing
High backpressure
The valve may be incorrectly plumbed; refer to
the instruction sheet supplied with the valve
The valve may be clogged; remove the valve
and clean it as described in the instruction sheet
supplied with the valve