Waveform of
some lead is
small in
amplitude
Equipment
connection
failure
Software was not turned off
normally after last use.
Please connect recorder and PC again
and restart the software.
PC has recovered from sleep
state, in sleep state, the power
will be cut off.
The Bluetooth adapter is
damaged.
Check and replace the Bluetooth
adapter.
The Bluetooth equipment is
not connected correctly.
Add the Bluetooth equipment again
according to the user manual.
Print failure or
wrong format
of report
There is no printer connected
or no printer driver installed
Please let professional technical
personnel to install the printer again,
and set the accuracy of printer as 600
dpi according to the user manual.
The interface
of software
can't be
displayed
entirely.
Resolution of display is too
low.
Adjust the resolution of display to
more than 1366*768 according to the
requirement described in the User
Manual.