Guidance and manufacturers declaration electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacturers declaration electromagnetic immunity
The CONTEC8000G is intended for use in the electromagnetic environment
specified below. The customer or the user of CONTEC8000G should assure that it is
used in such an environment.
Immunit
y test
IEC
60601 test
level
Com
pliance
level
Electromagnetic environment -
guidance
Electrosta
tic discharge
(ESD)
IEC
61000-4-2
6 kV
contact
8 kV
air
6
kV contact
8
kV air
Floors should be wood, concrete or
ceramic tile. If floor are covered with
synthetic material, the relative humidity
should be at least 30%.