(also suitable for use in newborn babies and pregnancy)
Use with GlucoRx HCT Glucose Test Strips,
GlucoRx HCT Ketone Test Strips and GlucoRx Lancets.
-6
5 15 25 35 45 55 65
-40
-20
0
40
20
60
Effect of haematocrit on accuracy
Haematocrit
% Error
Without 2+2
With 2+2
Haematocrit Correction Technology (HCT)
GlucoRx HCT Glucose test strips use advanced
GDH-FAD enzyme technology with 2+2 (AC and
DC) Bio Signal. AC signal works to measure the
haematocrit result, and DC measures glucose.
Meter then modulates glucose level according to
haematocrit value.
BGM vs. YSI VALUES BEFORE AND AFTER HCT
YSI (mmol/L)
Before 2+2
BGM (mmol/L )
After 2+2
10 25 50
50
25
10
0
4 Alarm
reminders
5 sec.
TIMER
5
Software
Features
Ketone testing & warning
Exceeds EN ISO 15197:2015 Accuracy criteria
Haematocrit Correction Technology
Measures Haematocrit (10 - 70% haematocrit
range)
4 Alarm reminders
5 second glucose results
requirements
Meal markers
Auto QC tagging
PC download facility; Diasend compatible.