29
Error Messages
MESSAGE WHAT IT MEANS WHAT TO DO
E-b
Appears when the batteries
are too low.
Replace the batteries
immediately.
E-C
Appears when the
wrong code strip is
inserted or other coding
errors.
Check if the code numbers on
the display and ketone foil
packet are the same. If the
problem persists, please
contact GlucoRx Customer care
on 01483 755133 for help.
E-2 Expired code strip.
Repeat the test with a new lot of
test strip.
E-U
Appears when a used test
strip is inserted.
Repeat with a new test strip.
E-t
Appears when ambient
temperature is above or
below system operation
range.
System operation range is
10°C to 40°C (50°F to 104°F).
Repeat the test after the meter
and test strip are in the above
temperature range.
E-0,
E-A,
E-E.
Problem with the meter.
Repeat the test with a new test
strip. If the meter still does not
work, please contact GlucoRx
Customer care on 01483
755133 for assistance.
E-F
Appears when test strip is
removed while counting
down, or insufcient blood
volume.
Review the instructions and
repeat test with a new strip.
If the problem persists, please
contact GlucoRx Customer care
on 01483 755133 for help.