27
4. After rinsing with water, allow the sticky sheet to air
dry for 3 to 5 minutes.
5. After drying completely, the sticky sheet can be
restored. Place it back in the carry case for use.
3~5min
NOTE: Never use any object to wipe off the sticky
sheet.
NOTE: The sticky sheet will become less sticky the
more you touch it, through transfer of your skin oils.
Important Control Solution Information
Use only GlucoRx control solution with your meter.
Do not use the control solution beyond the expiry date or 3 months
after first opening. Write the opening date on the control solution
vial and discard the remaining solution after 3 months.
It is recommended that the control solution test be done at room
temperature 20°C to 25°C (68°F to 77°F). Make sure your control
solution, meter, and test strips are at this specified temperature
range before testing.
Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
Store the control solution tightly closed at temperatures between
2°C to 30°C (35.6°F to 86°F). Do NOT freeze.