27
4. After rinsing with water, allow the sticky sheet to air
dry for 3 to 5 minutes.
5. After drying completely, the sticky sheet can be
restored. Place it back in the carry case for use.
3~5min
NOTE: Never use any object to wipe off the sticky
sheet.
NOTE: The sticky sheet will become less sticky the
more you touch it, through transfer of your skin oils.
Important Control Solution Information
● Use only GlucoRx control solution with your meter.
● Do not use the control solution beyond the expiry date or 3 months
after first opening. Write the opening date on the control solution
vial and discard the remaining solution after 3 months.
● It is recommended that the control solution test be done at room
temperature 20°C to 25°C (68°F to 77°F). Make sure your control
solution, meter, and test strips are at this specified temperature
range before testing.
● Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
● Store the control solution tightly closed at temperatures between
2°C to 30°C (35.6°F to 86°F). Do NOT freeze.