2 Recalibration
Click the ReCal. button in the Sensor Setup window of the particular sensor.
Example Turbidity:
[1] Radio buttons for selecting the gain factor
The recalibration is performed
at 1x and 10x and 100x
[2] Input fields for the reference values of the two reference fluids
[3] Display of the raw value of the current measurement in mV
[4] Display of the actual measured value
[5] Input fields of the measured raw values of the two reference fluids, the values can automatically be
transferred into the fields by clicking the buttons [6].
[6] Clicking a button transmits the measured raw value into the assigned field. Once values are entered in
both fields, the calibration coefficients are calculated.
[7] Setting of the digital filter (100% = no filter; 1% max filter) If the measured value is noisy, the filter can
improve the calibration.
[8] Display of the calibration coefficients (here named as „calibration factors“).
On operation with automatic range switching ( Automatic) you have to calibrate in all three areas.
Transmits the calibration on to the sensor.
Opens the window of the calibration history.