Settings and Measurement
(2) MENU key
Press the MENU key to switch through the AMP, ANNO, DATA, ALM, FILE, I/F, OTHR and INFO
setting menus. It can also be used to confirm details.
AMP ANNO DATA ALM FILE I/F OTHR INFO
qwertyui
qq
qq
q AMP Settings Window
AMP Menu Structure
Setting Selections available Setting method
Input Off, DC, TEMP ENTER→Select→ENTER
Range Voltage: 20, 50, 100, 200, 500 mV ENTER→Select→ENTER
1, 2, 5, 10, 20, 50, 1-5 V
Temperature: TC-K, TC-J, TC-T, TC-R, TC-E, TC-B,
TC-S, TC-N, TC-W, Pt100, JPt100
Filter Off, On ENTER→Select→ENTER
Function(EU) Off, On (effective when On has been selected) ENTER→Select→ENTER
Lower Setting – Settings
Value – Upper • Meas. Value (Upper/Lower)
ENTER→Set numeric value→ENTER
Unit • EU Value (Upper/Lower)
ENTER→Set numeric value→ENTER
• Dec pt ENTER→Select→ENTER
• Unit ENTER→Select→ENTER
• Select ENTER→Select→ENTER
Register ENTER
Color Current Color, Sel Color ENTER→Select→Register
Misc. Zero voltage adjustment Press ENTER to execute
• Perform Auto Zero ADJ.
• Reset Auto Zero ADJ.
[Zero point voltage value]
Misc
(only when ALL
Span All Settings ENTER→Set numeric value→
has been selected)
• Upper/Lower ENTER→Execute
Mode OFF, Revol., Counts, Inst. ENTER→Select→Register
Range Revol.: 500, 5k, 50k, 500k RPM/F.S. ENTER→Select→ENTER
Counts: 50k, 500k, 5M, 50M, 500M C/F.S.
Inst.: 50k, 500k, 5M, 50M C/F.S.
EU Function: Off, On ENTER→Select→ENTER
• Scaling settings ENTER→Select→Register
• Setting values (Use direction keys to select
• Unit settings numeric value)
• Unit
Slope H, L ENTER→Select→ENTER
Logic Off, On ENTER→Select→ENTER
Color specification
Specified color ENTER→Select→Register
EU (Scaling settings)
Pulse
Logic