Settings and Measurement
3-17
<Temperature Ranges>
<Humidity Range>
Filter .................................... Sets the filter status. Please set the filter to ON when there is likely to
be noise in the input. Filter operation is on a moving average basis.
Off, 2, 5, 10, 20, 40
EU (Scaling)........................ Scales the measured values and converts them to other units.
Function .............................. Sets the function to Off or On.
Upper, Lower, Unit .............. Sets the EU function's conversion value and unit.
If this message appears, follow the instructions by reducing the
number of digits to be output by one, or leaving the number of digits
as is and changing the EU value.
CAUTION
The Scaling operation is calculated using a ratio of the Meas. Value or
EU Output Value settings.
“++++/----” is displayed when the converted value cannot be processed
by GL800.
(a) Meas. Value
Specifies the numeric value to be converted. Set two points, the
Upper and Lower parameters.
Range Maximum SPAN Minimum SPAN (p-p) Measurement Range Minimum
Resolution
K -270 to +2000°C 50°C -200 to +1370°C
J -270 to +2000°C 50°C -200 to +1100°C
T -270 to +2000°C 50°C -200 to +400°C
R -270 to +2000°C 50°C 0 to +1600°C
E -270 to +2000°C 50°C -200 to +800°C
B -270 to +2000°C 50°C +600 to +1820°C 0.1°C
S -270 to +2000°C 50°C 0 to +1760°C
N -270 to +2000°C 50°C 0 to +1300°C
W -270 to +2000°C 50°C 0 to +2315°C
PT100 -270 to +2000°C 50°C -200 to +850°C
JPT100 -270 to +2000°C 50°C -200 to +500°C
PT1000 -270 to +2000°C 50°C -200 to +500°C
Range Maximum SPAN Minimum SPAN (p-p) Minimum Resolution
0 to +110% 1.0% 0.1%