Settings and Measurement
3-19
Logic and Pulse settings
Logic/Pulse Switching ......... Switches the digital input.
• Off: Digital input is disabled.
• Logic: Digital input is processed as logic signals.
• Pulse: Digital input is processed as pulse signals.
Filter............................... Sets the filter for digital input.
• Off: Disables filter circuit for digital input.
• On: Enables filter circuit for digital input.
Filter is approximately 30 Hz (-3 dB).
Pulse Input.................... Sets the pulse input method for each channel.
• Off: Pulse input is disabled.
• Revol.: Counts the number of pulses per second and
displays the value multiplied by 60 as rpm values.
• Counts: Displays the cumulative number of pulses for each
sampling interval from the start of measurement.
• Inst.:
Displays the number of pulses for each sampling interval.
Pulse Slope............ Sets the slope (direction) to count the number of pulses.
• H: Counts the rising pulse.
• L: Counts the falling pulse.
Pulse EU...................... Sets scales for pulse measurement values.
• Function: Sets On/Off of the Scaling function.
• Meas. Value: Specifies the numeric value to be converted.
• EU Value: Specifies output after conversion.
• Unit: Selects the converted unit, which can be specified as
a user-defined character string consisting of
alphanumerics. The Unit parameter can also be
specified by selecting the Select Unit setting.
• Select: Selects the type of engineering unit.
• Choose: Selects the converted unit. The Unit displayed here is
the type of unit selected by the Select Unit setting.