3-20
Settings and Measurement
DATA settings
This menu is used to specify capture-related items and calculations.
DATA Menu Structure
CAUTION
When you save files, create a folder and then save the files in the folder. Regardless of the remaining
capacity, if you try to save files in the root directory, due to file restrictions you may not be able to save files.
The displayed Capture Time may vary according to the sampling interval or number of capture channels.
Sampling Interval ................ Specifies the sampling interval for data capture.
The table below shows the number of measuring channels and sampling interval values that can
be set.
If data fluctuate due to noise, set the sampling interval to a value which enables the digital filter
function.
* "Number of Measuring Channels" is the number of channels in which input settings are NOT set to "OFF".
CAUTION
To use the digital filter function, you must set the AC power supply frequency accurately.
Follow the instructions on page 3-28 to ensure that the settings are accurate.
Captured data file name ..... Select the name of a file to which you want to save captured data.
Set either the main memory or USB device (option).
(See the file box on page 3-31.)
Setting Selections available
Record Settings
100, 200, 500ms 1, 2, 5, 10, 20, 30s
• Sampling Interval 1, 2, 5, 10, 20, 30min, 1h
File Name • File: Folder name, file name
• Name Type: Auto, User
• File Type: GBD, CSV
Statistical Calculation
Off, Average, Max, Min, Peak, RMS
• Function 1, 2
Number of Measuring Allowed Sampling Interval Sampling Interval which
Channels* enables Digital Filter
10 channels or less 100 ms or above 500 ms or above
11CH to 20CH 200ms or above 1s or above
21CH to 50CH 500ms or above 2s or above
51CH to 100CH 1s or above 5s or above
101CH to 200CH 2s or above 10s or above