Boston Scientific (User’s Manual Template 8.5in x 11in Global, 91091755 Rev/Ver. AA), Manual, MB, XPS, Global, 50623316-01A_pretrans
Black (K) ∆E ≤5.0
19
Process Black ΔE ≤5.0 / CMYK Image
13 TECHNICAL DATA
Table 1. Console Specifications
Laser Type Solid-state, frequency doubled
Wavelength 532 nm
Maximum Power Output at 532 nm Limited by fiber delivery device, up to 180 W
NominalOcularHazardDistance(NOHD) 67.1meters(MPE=10W/m
2
)
Repetition Rate Vaporization: Quasi-CW (23.6 kHz)
Coagulation: Modulated at ~12.5 Hz, ~25 % duty cycle
Maximum Aiming Beam Power 5 mW
OutputBeamDivergence ≥ 0.078 Radians
Electrical Requirements 200to240VAC,50/60Hz,≤20Amp,single-phase
Operating Temperature 50°Fto86°F(10°Cto30°C)
Operating Humidity 10 % to 90 %, non-condensing
Dimensions Width: Approximately 22 in (56 cm)
Depth:Approximately36in(91cm)
Height (screen down): Approximately 48 in (122 cm)
Height (screen up): Approximately 55 in (140 cm)
Weight Approximately 475 lb (215 kg)
AC Power Cable 15 ft (457 cm) maximum length, model 321203
AC power cable is a trademark of OLFLEX.
Footswitch Cable 13 ft (396 cm) maximum length, Cordset #997-G27-1
Footswitch cable is a trademark of LINEMASTER Switch Corporation.
Standards and Regulations • SafetyRegulationsconformswithEN/IEC60825
• EMFoutputconformswithIEC60601-1Generalrequirementsforsafety—Collateral
standard:Electromagneticcompatibility—Requirementsandtests
13.1 Safety Classifications and Electromagnetic Compatibility
Classification: Type of protection against electric shock: Class I equipment
Table 2. Electromagnetic Emissions
Guidance and Manufacturer’s Declaration — Electromagnetic Emissions
The GreenLight XPS™ Laser System is intended for use in the electromagnetic environment specified below. The customer or user of the GreenLight
XPS Laser System should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment — Guidance
RF Emissions
CISPR 11
Group 1 The GreenLight XPS Laser System uses RF energy only for its internal function. Therefore,
its RF emissions are low and are not likely to cause interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class A The GreenLight XPS Laser System is suitable for use in all establishments other than
domestic and may be used in domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
Fluctuations/Flicker
Emissions
IEC 61000-3-3
Section 5 of Standard
WARNING: Thisequipment/systemareintendedforusebyhealthcareprofessionalsonly.Thisequipment/systemmaycauseradiointerference
or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the
GreenLight XPS Laser System or shielding the location.