11. Troubleshooting
Please also refer to sensor documentation for additional troubleshooting guidelines.
Fault Cause Remedy
No water discharge
Inlet valve closed Open inlet valve complete
Inlet strainer blocked Unscrew cover of inlet strainer and clean the sieve
Flow detector or flow restrictor are
blocked
Clean flow detector or flow restrictor or replace complete part.
Water running from the
sampling point
Sampling point not correctly closed Close outlet fitting of sampling point
Power supply LED is off No power supply Connect power supply
Displayed system time is not
current or is not updated
every second
Internal clock not adjusted
Not updated every second
Adjust clock in menu Setup
Restart CU 382
Error messages or error
symbol displayed?
Sensor communicates an error
Go to Parameter settings / Monitor and check status (see
sensor manual for status information)
Tubes or connections are
leaking
Fittings are not tight
Try to tighten fittings. Be careful when using heavy tools for
tightening. Tubes, fittings and flow cell can be destroyed if too
much force is used!
Flow detector shows no flow
Inlet valve is closed
Inlet strainer is blocked
Open inlet valve
Clean inlet strainer
Water is not visible within the
tubes
Inlet valve is closed
Inlet strainer is blocked
Open inlet valve
Clean inlet strainer
Water is not visible within the
tubes
If pressure on the inlet is very high, air
can come out after the flow restrictor
Reduce pressure on inlet
Parameter readings are
fluctuating
Measuring electrode is contaminated
or faulty
Inlet strainer is blocked
Air bubbles inside flow cell
Exchange electrolyte or membrane cap (see sensor
manual for more information)
Clean inlet strainer
Reduce pressure in inlet
Parameter reading are not
plausibly
Measuring electrode is contaminated
or faulty
Inlet strainer is blocked
Air bubbles inside flow cell
Exchange electrolyte or membrane cap (see sensor
manual for more information)
Clean inlet strainer
Reduce pressure in inlet
Readings are not within the
specified and calibrated
measuring range
Sensor / DID was powered on recently
Concentration of measurement
substance is to high
It may take up to an hour before a sensor provides correct
values
Changing to sensor with a wider measurement range