D-0120695 Rev D 2022-06 Page 14
levels. Press the key to confirm the selection or the key to cancel. (Note that 100dBHL at
4000Hz is not available).
REFLEX POLARITY
Use the and keys to choose whether the reflex traces are displayed as ascending (UP) or
descending (DOWN). Press the key to confirm the selection or the key to cancel.
REFLEX FILTER
Use the keys to choose either 2Hz or 1.5Hz. The default of 2Hz is suitable for most circumstances.
However, if a smoother reflex plot is required for better interpretation 1.5Hz may be chosen.
Press the key to confirm the selection or the key to cancel.
SET TIME/DATE
Use the keys to enter the values for the date and time. Use the and keys to change the
values. Press the key to confirm and save the selection or the key to cancel.
POWER OFF DELAY
The GSI Allegro will switch off automatically if no key is pressed for a specified duration. Use the
and keys to change this duration between 90 and 180 seconds and press the key to
confirm and save the selection or the key to cancel.
LCD CONTRAST
Use the and keys to change the contrast of the LCD screen; press the key to confirm and
save the selection or the key to cancel.
REPORT CAL DATE
The printout of the test results may include date of the instrument’s calibration. Use the and
keys to select if the calibration date is printed or hidden. Press the key to confirm and save
the selection or the key to cancel.
SET DATE FORMAT
The GSI Allegro supports two different date formats. Use the and keys to select either
DD/MM/YY or MM/DD/YY and press the key to confirm and save the selection or the key
to cancel.
HOSPITAL NAME
The printout of the test results may include the hospital name (up to 19 characters). To enter
the hospital name use the and and and keys to select the letter then press and briefly
hold the key to confirm. To delete the last letter briefly hold the key. Once the name has
been entered highlight the # key then press and briefly hold the key to save the name.
Highlight the # key then press and briefly hold the key to cancel.