Allegro Tympanometer
User Manual
D-0120695 Rev D 2022-06 Page 23
SAVE RESULTS (Save the results in the internal database)
VIEW TEST (Review the results as described above)
MAIN MENU (Return to the main menu)
The results of the last test performed remain available even if the Allegro has been turned off.
To view these results select VIEW THE LAST TEST from the main menu. After selecting the
required ear, the tympanogram will be displayed. It will then be possible to view the results and
select the PROCESS RESULTS menu as if the test had just been completed.
NOTE: Results of the last test will be erased as soon as a new test is started. Test results
should be saved to the internal database or printed to ensure that data is not lost.
ERROR MESSAGES
The following error messages may be seen during the test sequence.
Message Displayed
Indicator
Status
Likely Cause(s)
WITHDRAW PROBE
Yellow
Flashing
The probe has been moved during measurement.
Re-insert the probe to repeat the test.
Volume outside range
WITHDRAW PROBE
Yellow
Flashing
The ear canal volume is above the 5ml. This
message can also occur when the probe is not
properly inserted into the ear.
Blocked ear
WITHDRAW PROBE
Green
Flashing
The ear canal volume is below 0.1ml. Check that
the probe is correctly inserted into the ear. Also
check that the probe is not blocked.
INSERT PROBE
Yellow
Flashing
The seal was lost. Reinsert the probe to repeat
the test.