ERROR MESSAGES & FAULT CONDITIONS
CAUTION
If a fault condition cannot be cleared, the operator is cautioned against repeatedly
starting the instrument. In some fault conditions the internal pump may
progressively advance towards the end of its travel to clear the fault. If the end of
travel is reached in such conditions the instrument may lock up and become un-usable.
If difficulties resolving fault conditions occur the equipment distributor should be consulted.
Please ensure the probe is not
blocked or obstructed
Examine the probe tip for blockages. If necessary,
remove it and clean or replace it. If the problem
persists, contact your GSI service center.
Unknown pump fault. Restart the
unit. If problem persists, contact GSI
WARNING! CALIBRATION EXPIRED.
Recalibration needed before further
tests are performed
The current date is later than the next calibration
date. Check that the clock is set to the correct date. If
so, arrange for the instrument to be recalibrated.
Tests can still be performed.
Recharge the batteries before
performing tests
Recharge the batteries immediately
Other than after the specified power off delay, the
Allegro may turn off because the internal batteries
are spent. To replace the batteries, contact your GSI
service center.
Cannot determine pump direction. If
problem persists, contact GSI
Pump fault. If the fault persists, contact your GSI
service center.