Select Waveform Update Mode
Background
Normal
Updates the whole displayed waveform at
once. Automatically selected when the
timebase (sampling rate) is fast.
Timebase
≤50ms/div (≥50kSa/s)
Trigger
all modes
Roll Mode
Updates and moves the waveform
gradually from the right side of the
display to the left. Automatically selected
when the timebase (sampling rate) is slow.
Timebase
100ms/div
(≤25kSa/S) 350MHz models
(≤20kSa/S) 500MHz models
Trigger
all modes
Roll
Select Roll Mode
Manually
1. Press the Trigger Menu key.
Menu
2. Press Mode from the bottom menu
and select Auto (Untriggered Roll)
from the side menu.