MEASUREMENT
93
Start Go-NoGo
Press Enable to start the Go-NoGo test.
The Enable button will change to
Disable. Pressing Disable will stop the
Go-NoGo test and toggle the button
back to Enable.
Exit the
Application
To exit the application, press Break.
Using the Go-
NoGo Output
To output the Go-NoGo results to an
external device, the Go-NoGo rear
panel terminal (open collector) can be
used. The Go-NoGo terminal will
output a positive pulse each time a
NoGo violation has occurred for a
minimum of 10us. The voltage of the
pulse depends on the external pull-up
voltage.
The line out socket can also be used to
connect directly to a buzzer.
Line Out
Timing Diagram
Output timing
Enable
Go-NoGo
Output
500us
Waveform
acquisition
When a No-Go violation occurs the Go-NoGo output
will go from a low level to a high level for a minimum
of 500us.

Table of Contents