ANALOG CONTROL
145
Isolated External Voltage Control of Voltage Output
Background
Voltage control of the voltage output uses the
isolated voltage option (PSU-ISO-V). A voltage of
0~5V or 0~10V is used to control the full scale
voltage of the instrument, where:
For 0~5V: Output voltage = full scale voltage ×
(external voltage/5)
For 0~10V: Output voltage = full scale voltage ×
(external voltage/10)
Connection
When connecting the external voltage source to the
isolated voltage option, use shielded or twisted
paired wiring.
PSUEXT-V
Isolated
Voltage
Option
4
2
Output
Terminal
2 core shielded
wire or twisted
pair
1or8
Pin4(GND_ISO) → EXT-V (-)
Pin2(+VPROG_ISO) → EXT-V (+)
Pin 1 or 8 (Shield)Wire shield