EasyManua.ls Logo

GW Instek PSW 30-108 - Page 103

GW Instek PSW 30-108
187 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CONFIGURATION
103
Show Version
F-89
0, 1 = PSW version
2, 3 = PSW build year
4, 5 = PSW build month/day
6, 7 = Keyboard CPLD version
8, 9 = Analog-Control CPLD version
A, B = Reserved
C, D = Kernel build year
E, F = Kernel build month/day
G, H = Test command version
I, J = Test command build year
K, L = Test command build
month/day
M, N = USB Driver version.
Power On Configuration Settings*
CV Control
F-90
0 = Panel control (local)
1 = External voltage control
2 = External resistance control
(Ext-R 10k = Vo, max)
3 = External resistance control
(Ext-R 10k = 0)
CC Control
F-91
0 = Panel control (local)
1 = External voltage control
2 = External resistance control
(Ext-R 10k = Io,max)
3 = External resistance control
(Ext-R 10k = 0)
Power-ON Output
F-92
0 = OFF at startup
1 = ON at startup
T001 ~ T010 = Run test script TXX at
start up
Master/Slave
F-93
0 = Master/Local
1 = Master/Parallel1
2 = Master/Parallel2
3 = Slave/Parallel
4 = Slave/Series (Only 30V, 80V, 160V
models)
External Out Logic
F-94
0 = High ON, 1 = Low ON
Power Switch trip
F-95
0 = Enable , 1 = Disable
Calibration Settings*
Calibration
F-00
0000 ~ 9999
www.GlobalTestSupply.com