Analog signal common. Connected to the sense- 
terminal when remote sense is used. Connected to 
the negative output terminal when remote sense is 
not used. 
Common for status signals 18, 19, 20, 21 and 22. 
Turns on when CV mode is active. (photo coupled 
open collector output) 
Turns on when CC mode is active. (photo coupled 
open collector output) 
Turns on when any of the protection modes are 
tripped (OVP, OCP) or if a shutdown signal is 
input. (photo coupled open collector output) 
Turns on when the output has been turned on. 
(photo coupled open collector output) 
Turns on when the power switch is turned off.  
Turns the output on/off when (default setting) a 
low TTL signal is applied. Internally, the circuit is 
pulled up to +5V with 10kΩ resistance. 
Series slave input during master-slave series 
operation. (30V/80V/160V models only) 
External Voltage Control of Voltage Output 
 
External voltage control of the voltage output is 
accomplished using the MIL-26 connector on 
the rear panel. A voltage of 0~10V is used to 
control the full scale voltage of the instrument, 
where: 
Output voltage = full scale voltage × (external 
voltage/10) 