REMOTE CONTROL 
31 
Configure GPIB Interface  
To use GPIB, the optional GPIB to USB (GUG-001) adapter must be 
used. The GPIB to USB adapter must be connected before the PSW 
is turned on. Only one GPIB address can be used at a time. 
 
1.  Ensure the PSW is off before proceeding. 
2.  Connect the USB cable from the rear panel  
USB B port on the PSW to the USB A port on 
the GPIB to USB adapter. 
3.  Connect a GPIB cable from a GPIB controller to 
the GPIB port on the adapter. 
Type B plug for  
PSW series
Type A plug
From 
computer
GUG-001
5.  Press the Function key to enter the 
Normal configuration settings.  
 
      Set the following GPIB settings: 
Set the rear panel USB port to 
GPIB-USB (GUG-001) 
Set the GPIB address (0~30) 
  Maximum 15 devices altogether, 20m cable 
length, 2m between each device 
  Unique address assigned to each device 
  At least 2/3 of the devices turned On 
  No loop or parallel connection