What happens if no stir bar is installed in Hach Measuring Instruments?
- Jjames13Sep 12, 2025
If no stir bar is installed or more than one stir bar is installed in your Hach Measuring Instruments: * Make sure that a stir bar is installed. * Use a paper clip to retrieve the stir bar from the colorimeter and make sure that multiple stir bars are not installed. * If the stir bar does not operate, mix the sample manually after the reagents have been added. Use the end of a wooden Q-tip and mix the reagents approximately for 10 seconds. If the instrument gets a reading, the mixing is not sufficient. Replace the stir coil. Additional information is available on the manufacturer's website.