How to fix a pinched sample line in the pinch block of Hach 5410001 Measuring Instruments?
- SStephen HunterSep 3, 2025
If the sample line is pinched off in the pinch block of your Hach Measuring Instruments, follow these steps: * Pull the sample line from the colorimeter, being aware that water in the line may leak out. Examine and adjust the flow rate to 200 mL/minute while the line is pinched, ensuring a steady water stream. * Make sure the pressure plate of the pinch block isn't overtightened. * Inspect the back of the pinch blocks for grooves.