Table of Contents
1 Specifications on page 3
2 General information on page 4
3 Installation on page 7
4 User interface and navigation on page 16
5 Startup on page 17
6 Operation on page 18
7 Maintenance on page 20
8 Troubleshooting on page 27
9 Replacement parts on page 30
Section 1 Specifications
Specifications are subject to change without notice.
Specification Details
Dimensions (W x H x D) 34.11 x 47.50 x 17.93 cm (13.43 x 18.70 x 7.06 in.)
Enclosure IP62 rating when the door is latched (indoor use)
Weight 11.3 kg (25 lb)
Installation category II
Pollution degree 2
Protection class I
Altitude 2000 m (6560 ft)
Display LCD, 3.5-digit measurement readout and six-character alphanumeric scrolling text line
Mounting Wall mount
Power requirements 100–115/230 VAC; 90 VA, 50/60 Hz; 2.5 A fuse
Power connection 12 to 18 AWG wire
Operating temperature 5 to 40 °C (41 to 104 °F)
Storage temperature –40 to 60 °C (–40 to 140 °F)
Humidity 90% at 40 °C (90% at 104 °F)
Air purge (optional) 0.1 CFM instrument-quality air, ¼-inch O.D. tubing
Alarm relay outputs Two unpowered SPDT relays, each rated at 5 A resistive, 240 VAC maximum.
Alarm connection 12 to 18 AWG wire
Recorder output One isolated recorder output, 4–20 mA (can be adjusted to 0–20 mA), maximum
impedance 500 ohms
Recorder connection 12 to 18 AWG wire
Maximum reagent usage 0.5 L per month (for each of the two standards and one reagent)
Reagent containers High-density polyethylene 0.5 L bottles contained inside the analyzer enclosure and
vented externally.
Measurement range 0.1 to 10 mg/L fluoride
Accuracy/precision ±10% or ±0.10 ppm; ±7% or ±0.07 ppm
Cycle limit 4.2 minutes
Calibration Two-point log/linear with 0.5 mg/L and 5.0 mg/L fluoride standards
Detection limit 0.10 ppm
Certifications CE. Listed to UL and CSA safety standards by ETL.
English 3

Related product manuals