EasyManuals Logo

Hach LPV445.99.02110 Service Manual

Hach LPV445.99.02110
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
D A N G E R
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
The generic steps to do a test follow.
To do a test for a specific parameter (e.g., chlorine), download the test procedure from
the manufacturer's website. Refer to Download a test procedure on page 11.
1. Push to select the applicable measurement range (e.g., LR or HR).
2. Prepare the blank. Refer to the test procedure.
3. Clean the sample cell with a no-lint cloth.
4. Insert the blank sample cell into the cell holder. Make sure to install the blank sample
cell in the correct and consistent orientation so that the results are more repeatable
and precise. Refer to Figure 5.
5. Install the instrument cap over the cell holder. Refer to Figure 6.
6. Push to set the instrument zero.
7. Remove the blank sample cell.
8. Prepare the sample. Refer to the test procedure.
9. Clean the sample cell with a no-lint cloth.
10. Insert the sample cell into the cell holder. Make sure to install the sample cell in the
correct and consistent orientation so that the results are more repeatable and precise.
Refer to Figure 5.
11. Install the instrument cap over the cell holder. Refer to Figure 6.
12. Push . The display shows the results in concentration units or absorbance.
Note: The result flashes if the result is less or more than the instrument range.
13. Remove the sample cell from the cell holder.
14. Immediately empty and rinse the sample cell. Rinse the sample cell and cap three
times with deionized water (or distilled water).
Note: As an alternative, use tap water to rinse the sample cell if the samples measured have a
higher concentration than the tap water.
Startup
10

Other manuals for Hach LPV445.99.02110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach LPV445.99.02110 and is the answer not in the manual?

Hach LPV445.99.02110 Specifications

General IconGeneral
BrandHach
ModelLPV445.99.02110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals