Figure 11 Get the instrument log
Figure 12 Instrument logger
Category Code Value Short description Detailed description
error 15 Error during software update [DR300]SUBKEY_UPDATE_FAILED_MISC = 0U
SUBKEY_UPDATE_FAILED_SIGN_ERR =1U
SUBKEY_UPDATE_FAILED_CRC_ERROR = 2U
SUBKEY_UPDATE_FAILED_CANCELED = 3U
[none] Error type
error 17 Warning: battery capacity is low [DR300]Warning: battery capacity is < 2%
[V] Battery Voltage
[%] Battery Capacity
[°C] Temperature
[V] Hardware ID Volt
[V] Module ID Volt
standard
message
23 Status: serial number was changed [Default]No further information for this event.
[none] new serial number
[none] previous serial number
standard
message
26 Status: start SW-Update [Default]No further information for this event.
Troubleshooting