8. On the clean polishing cloth (2934) do the steps that follow:
a. Add a few drops of water.
b. Spread a little of the polishing powder (29331) to form a grey, milky liquid.
9. Screw the polishing tool (DG33629) onto the top of the sensor.
10. Polish the electrodes by moving the sensor face in a circular direction against the liquid in the
polishing cloth for about 30 seconds.
11. Remove the polishing tool from the sensor. Remove any polish deposits by rinsing the sensor
head under a tap for 30 seconds, aiming the jet of water directly onto the sensor head.
Note: If the sensor is used in high level hydrogen sample, do not perform the following steps but continue at
step 19.
12. Remove the plastic base from the bottom of the sensor and connect the sensor to the sensor
cleaning and regeneration center (32301).
16
English

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach ORBISPHERE Series and is the answer not in the manual?

Hach ORBISPHERE Series Specifications

General IconGeneral
CategoryAccessories
ManufacturerHach
ModelORBISPHERE Series
Power SupplyBattery or external power supply (depending on model)
CommunicationAnalog output, digital communication (e.g., Modbus, Profibus, HART) depending on model
DimensionsVaries depending on model
WeightVaries depending on model
Measurement Range0 ppb - 50 ppm (depending on sensor)
Accuracy±1 ppb or ±1% of reading (depending on sensor)
Operating Temperature0 to 50°C (32 to 122°F) (depending on sensor and instrument)

Related product manuals