Transmitter 9135 - pH/redox measurement
Page 50
S1 & S2 RELAYS
− Proportionality coefficient : slope of regulation.
COEF.XP1/2 : XXX – In %
− Value of the pulse width (0,1 < T < 0,7) :
T1/2 : XXX – In %
− Minimum number of pulses per minute. Program
this value according to the control system
associated to the relay :
MIN : XXX – In i/min
− Maximum number of pulses per minute. Program
this value according to the control system
associated to the relay.
MAX : XXX – In i/min
MANUAL MODE
− Force relay S1 or S2. Pressing the function key
associated to the relay, the relay is activated till
you release the key. The corresponding icon is
displayed according to the relay state.
PULSE CONTROL :
The pulse period is constant and programmable (3
to 60 s). It is the cycle quotient (α) which varies
according to the sample pH value.
Off
On
"
?
"
?
"
?
S3
T T T
COEF.XP1 : 100%
T1 : 2.00 s
MIN : 001 i/min
MAX : 200 i/min
RELAY S1
Select
COEF.XP2 : 100%
T2 : 2.00 s
MIN : 001 i/min
MAX : 200 i/min
RELAY S2
Select
7.00