Transmitter 9135 – pH/redox measurement
Page 51
CONTROLLER
− Regulation form select
MODE : Pulse
S1 & S2 RELAYS
− Value of the pulse period (3 < T < 60) :
T1/2 : XXX In s
− Minimum activation time of the relay :
S1/2 MIN : XXX In s
− Value of the power factor. In the case of a non-
linear calculation of the control value, modify the
exponent value. This modification allows a
greater correction of the measurement when it is
far from the setpoint.
"PULSE/FREQUENCY" CONTROL OR
"FREQUENCY/PULSE" CONTROL
The controller may be programmed in a mixed
mode. Use the Mode command for the
programming. See paragraphs concerning the
frequency control and the pulse control to program
the relay S1 and S2.
CNTRLLR : Yes
MODE : Imp.
POINT : 7.00 pH
NEUT. Z : 2.0 pH
INERTIA : 000 s
CTL TIME : 20 mn
RELAY S1
RELAY S2
MANUAL MODE
CONTROLLER
Select