Table 9 Measurement (continued)
Error/Warning Description Possible solution
Equiv. point not found
The titration fails to determine the
equivalence point
Refer to the application note for best practices
and do the titration again
Adjust the equivalence point detection value for
the application (refer to Change application
parameters on page 63)
Contact technical support
Out of range
The computed result is outside the accepted
limits
Refer to the application note for best practices
and do the titration again
Adjust the result minimum and maximum values
for the application (refer to Change application
parameters on page 63)
Equation calc. error
An equation variable is unknown (e.g. the
ordinate of the equivalence point is unknown
or the volume of the equivalence point is
unknown)
Make sure that all the equation variables are
available (refer to the application note for the
equation description and values)
A required electrode is not
connected!
Application {application
name}
Electrode: Type: {type}:
Name: {name}:
The applicable IntelliCAL electrode type for
the application is not connected.
Connect the IntelliCAL electrode specified in the
application note.
The "Sensor name" set in the legacy adapter
settings does not agree with application
settings.
Change the "Sensor name" setting in the legacy
adapter setting to agree with the sensor name
specified in the application note.
Electrode calibration has
expired
A titration or titrant calibration has been done
with an expired sensor calibration date
Calibrate the sensor and if necessary adjust the
calibration frequency (refer to the application
note for the recommended frequency)
Titrant calibration has
expired
A titration has been done with an expired
titrant calibration date
Calibrate the titrant and if necessary adjust the
calibration frequency (refer to the application
note for the recommended frequency)
The titrant used by an
application has been
modified and this
application became
incompatible with
applications already
installed. Please go to
application management to
solve the issue
The application data has been changed and
the application cannot be used
Remove the application or change the titrant to
a titrant that is already installed
System error
A system error causes the titration to stop
(e.g. a sensor has been disconnected or the
volume of reagent is too low)
Make sure that all the system elements are
connected and working correctly (refer to
Maintenance menu on page 42)
Max. vol. reached
The titration is stopped because the supplied
volume of reagent is more than the maximum
quantity
Refer to the application note for best practices
and do the titration again
Adjust the maximum volume stop point for the
application (refer to Change application
parameters on page 63)
No titrant information There is no description of the titrant Install the application again
Troubleshooting