Appendix A Change application parameters
N O T I C E
The applications installed on the instrument have been pre-defined to optimize the measurement
process. Changing these default application parameters will have an effect on the measurement
process and measurement results. Only qualified personnel should change these parameters or
use the instrument in Expert mode. Refer to the application note before changing any
parameters.
N O T I C E
In Basic mode changes to parameters are secure. In Expert mode changes to parameters are
not secure and can cause the application to run incorrectly.
N O T I C E
The Active parameter is available to change in some parameter groups. It specifies if the method
should be used during the analysis. Changing this from Yes to No means that the method will not
be used which will have an effect on the measurement process and measurement results.
N O T I C E
The Hide result parameter (e.g. R2 hide, R3 hide, etc.) is available to change in some parameter
groups. It specifies if the result is shown on the display. Changing this from Yes to No means that
the calculated measurement result will not be shown. This will not have an effect on the
measurement process and measurement results.
The parameters available for change can be different for each instrument configuration
and the application being edited. The parameters marked with an "x" are available for
change.
Table 18 Application
Parameter Description Basic mode Expert mode
Name Name of the application x x
Advisable syringe The application has been designed with this type of syringe
Table 19 Sample
Parameter Description Basic mode Expert mode
Name Name of the sample x x
Amount Amount of the sample x x
Unit Unit of the sample x
Min. amount Minimum permitted sample amount x
Max. amount Maximum permitted sample amount x
Resolution Number of decimals to show for the sample amount x
Table 20 QC
Parameter Description Basic mode Expert mode
Name Name of the QC x x
Table 21 Blank
Parameter Description Basic mode Expert mode
Description Description of the blank x x
63