Table 22 Electrode
Parameter Description Basic mode Expert mode
Type Type of measurement electrode
Part of name Part of the name used to identify matching electrodes if the
preferred electrode is not connected
Recommended electrode Recommended electrode to be used for this analysis x
Calibration frequency Recommended frequency of calibration (0 = no reminder) x x
Stability criterion Refer to the working procedure for the recommended value x
Max. stability time Maximum time limit for the stability check x
Stirring speed Speed of the stirrer during electrode calibration x x
pH electrode parameters
Stirring duration Stirring duration before buffer measurement (0 = buffer
measurement with stirring)
x
Max. temp. variation Maximum permitted temperature variation during calibration
Min. slope Minimum permitted slope limit for the calibration result x
Max. slope Maximum permitted slope limit for the calibration result x
Min. offset Minimum permitted offset limit for the calibration result x
Max. offset Maximum permitted offset limit for the calibration result x
Calibration mode Set the electrode calibration mode x
Buffer set List of available buffers for electrode calibration x x
Number fixed buffers The number of fixed buffers to use
Fixed buffer #n Buffer to be used during calibration x
Conductivity electrode parameters
Calibration mode Set the electrode calibration mode x
Standard Standard used for electrode calibration x x
Min. cell constant Minimum permitted cell constant for the calibration result x
Max. cell constant Maximum permitted cell constant for the calibration result x
Temp. compensation Follow Natural Water Standards ISO/DIS_7888 x
Linear temp. coeff. Coefficient for linear temperature compensation mode x
Reference temp. Reference temperature for linear temperature compensation mode x
PtPt electrode parameters
Working mode Specify if the probe is used in imposed voltage/current (AC or DC) x
Current (sample analysis) Set the value of the imposed current during sample analysis x
Current (titrant calibration) Set the value of the imposed current during titrant calibration x
Voltage (sample analysis) Set the value of the imposed voltage during sample analysis x
Voltage (titrant calibration) Set the value of the imposed voltage during titrant calibration x
Change application parameters