Table 23 Titrant
Parameter Description Basic
mode
Expert mode
Name Name of the titrant x
Titrant concentration Nominal concentration of the titrant x
Resolution Number of decimals of the nominal concentration of the titrant x
Unit Unit of the titrant x
Real concentration Real concentration of the titrant used for result calculation x x
Resolution Number of decimals of the real concentration of the titrant x
Location The syringe that contains the titrant
Titrant calibration
Active Select No to skip this method during titrant calibration x x
Calibration frequency Recommended frequency of titrant calibration (0 = no reminder) x x
Stirring speed Speed of the stirrer during titrant calibration x x
Incremental addition
Predose type Type of the predose (in volume or in ordinate) x
Predose volume Titrant volume added at the beginning of the titration x
Predose ordinate Ordinate measurement to reach to stop the predose x
Predose speed Speed in mL/min during the predose in ordinate x
Delay Stirring time before the titration starts x
Max. vol. stop point Stops the titration when this volume is reached x
Ordinate stop point Stops the titration when this ordinate is reached x
Stop on last EQP Stops the titration on the last equivalent point detection x
Min. ordinate Contact technical support before changing. The minimum ordinate value
is used in the regulation algorithm during the dynamic addition
x
Max. ordinate Contact technical support before changing. The maximum ordinate value
is used in the regulation algorithm during the dynamic addition
x
Min. angle Contact technical support before changing. The minimum tangent angle
value is used in the regulation algorithm during the dynamic addition
x
Max. angle Contact technical support before changing. The maximum tangent angle
value is used in the regulation algorithm during the dynamic addition
x
Min. stability time Minimum time after an increment addition x
Max. stability time Maximum time after an increment addition x
Stability criterion Stability criterion after an increment addition x
Increment size Increment size when the addition is monotonic x
Min. increment size Minimum increment volume when the addition is dynamic x
Max. increment size Maximum increment volume when the addition is dynamic x
EP detection
EP1 ordinate Ordinate of this end point x
Change application parameters
65